Instrument Calibration

Instrument Calibration

Calibration is increasingly important as international requirements become stricter. Compliance and qualification testing need traceable proof of the performance of the equipment used for testing and measurement.

At Pangolim we offer a number of calibration services conducted at our calibration laboratory. We can also offer a on-site calibration service to customers that have large amounts of equipment minimising interruptions to operations.

Our calibration services include:

Electrical

Equipment TypeOur CapabilityReference Standard Accuracy
Digital Multi Meters (All types)
Including:
Loop calibrators RCD testers
Temperature by electrical
simulation
DCV -1000 to 1000
ACV 0 to 1000V
ACI 0 to 30A
DCI 0 to 30
Resistance. 0 to 1G
Freq. 0 to 10 MHz
0 to 1200°C
±8ppm
±0.015% (20mV -1025V,10Hz1MH)
0.04% (20uA-30A, 10Hz-30kHz)
±50ppm (0- 1025V)
±18ppm (Resistance by simulation)
±0.09°C
Insulation TestersUp to 1000G (1T )
Insulation Voltage: Up to
15KV
±100ppm
Clamp Meters2500 Amps on 10, 20
and 50 Turn coil
±8ppm
±0.015% (20mV - 1025V,10Hz - 1MH)
0.04%(20uA-30A, 10Hz-30kHz)
0.04% (20uA-30A, 10Hz-30kHz)
±50ppm
Thermocouple (simulation)-200 to 1700°C±0.09°C
Power Supply0 to 30 A
0 to 60 V
±8ppm
±0.015% (20mV - 1025V,10Hz - 1MH)
0.04% (20uA-30A, 10Hz-30kHz)
±50ppm (0- 1025V)
Optical Tachometer1 to 60,000 rph±200ppm
Power Quality AnalyzerDCV -1000 to 1000
ACV 0 to 1000V
ACI 0 to 30A
DCI 0 to 30
Resis. 0 to 10 M
Freq. 0 to 10 MHz
±8ppm
±0.015% (20mV - 1025V,10Hz - 1MH)
0.04% (20uA-30A, 10Hz-30kHz)
±50ppm (0- 1025V)
±18ppm (Resistance by simulation)
±1ppm (Frequency, all forms)
Scope Meter/OscilloscopesFrequency: 25MHz
Amplitude: 1000v/min
Voltage:
DCV 0 to 1000V
ACV 0 to 1000V
ACI 0 to 30A
DCI 0 to 30
±1ppm
±200ppm
±8ppm
±0.015% (20mV - 1025V,10Hz - 1MH)
0.04% (20uA-30A, 10Hz-30kHz)
±50ppm (0- 1025V)
±15ppm

Pressure

Equipment TypeOur CapabilityReference Standard Accuracy
Pressure gauge0 to 1000 bar0.05%FS
Pressure transmitters0 to 1000 bar0.05%FS
Pressure Modules0 to 1000 bar0.05%FS
Pressure Calibrators0 to 2000mbar
0 to 100 bar
0.02%FS
0.05%FS

Temperature

Equipment TypeOur CapabilityReference Standard Accuracy
LIG / MIG
Thermometer
-25 to 140°C0.05%FS
Digital Thermometer-25 to 140°C0.05%FS
Temperature Block
Calibrators
-35 to 650°C0.05%FS
Thermocouple (measure, all
types)
0 to 1200°C0.02%FS
0.05%FS
IR Thermometer0 to 140°C±0.03°C
Temperature transmitters0 to 100°C±0.01°C
Probes with readout
indicators
0 to 100°C±0.01°C

Why Choose Pangolim for Calibration Services?

At Pangolim we provide ourselves on providing a full, detailed, and accessible calibration service to our clients.
Our comprehensive calibration services include the following:-

  • Quick turnaround
  • Express turnaround services
  • Digital certificates
  • On-site surveys
  • Equipment calibrated to available manufacturer specifications
  • Comprehensive technical support and advice on all calibration requirements
  • Computerised automatic equipment recall prior to calibration due date
  • Calibration labelling and security validity labelling

To find out more about our calibration services, please get in touch with a member of the Pangolim team today, and we’ll be more than happy to answer any questions you may have regarding our services and how we can help you.